In the Altitude Test Chamber composite fault repair cases, the technical team through the three-level diagnostic method to accurately locate the PLC hardware failure and mechanical seal failure two core issues. The trouble shooting process of this Altitude Test Chamber provides a systematic maintenance reference for the industry:



A model of Altitude Test Chamber in the simulation of 15.2km altitude conditions, the compressor overload alarm and low pressure alarm triggered at the same time. The technical team used "refrigerant system → operation logic → electrical control → sealing system" four-step investigation method:
1. Refrigerant system verification: measured dual compressor equilibrium pressure stabilized at 150PSI, to rule out the possibility of Altitude Test Chamber refrigerant leakage;
2. Operation logic abnormalities: found that only the C2 compressor to participate in the control of temperature, C1 continues to dissipate the heat for the C2 resulting in low-pressure alarms;
3. PLC hardware failure: Detection of PLC output module no signal, confirm the Altitude Test Chamber control circuit relay aging damage;
4. Sealing system failure: motor fixation block screws loose, door latch slack leads to box leakage.
control circuit upgrade
Mechanical seal reinforcement

The case reveals the typical characteristics of compound failure of Altitude Test Chamber: more than 80% of the failure originates from the coupling failure of control system and mechanical structure. The technical team recommends the establishment of a three-dimensional prevention mechanism - daily operating parameters comparison, monthly vibration spectrum analysis, semi-annual sealing performance verification, which can reduce the risk of unplanned downtime by more than 60%.